dc.description.abstractA sensitivity improvement for arsenic determination in the presence of copper by cathodic stripping voltammetry, using a high concentration of hydrochloric acid (3 mol/L) and high cell temperature (40 degrees C), is described. The optimum conditions were established, and LOD and LOQ for determination of As(III) were calculated from the calibration curve to be 7 ng/L and 22 ng/L, respectively. The accuracy of the method was checked by analyzing a certified reference material (LGC 6010-Hard Drinking Water). The method was applied to drinking water samples in Balikesir city of Turkey.en_US
dc.language.isoengen_US
dc.publisherParlar Scientific Publications (P S P)en_US
dc.rightsinfo:eu-repo/semantics/openAccessen_US
dc.subjectArsenicen_US
dc.subjectCathodic Stripping Voltammetryen_US
dc.subjectDrinking Wateren_US
dc.titleDetermination of arsenic by cathodic stripping voltammetry: Sensitivity improvement using high concentration of hydrochloric acid and high temperatureen_US
dc.typearticleen_US
dc.relation.journalFresenius Environmental Bulletinen_US
dc.contributor.departmentFen Edebiyat Fakültesien_US
dc.contributor.authorID0000-0002-5527-6472en_US
dc.identifier.volume15en_US
dc.identifier.issue6en_US
dc.identifier.startpage542en_US
dc.identifier.endpage547en_US
dc.relation.publicationcategoryMakale - Uluslararası Hakemli Dergi - Kurum Öğretim Elemanıen_US


Bu öğenin dosyaları:

Thumbnail

Bu öğe aşağıdaki koleksiyon(lar)da görünmektedir.

Basit öğe kaydını göster